| Hazard | Source | Initial Mitigation Direction |
|---|---|---|
| Patient leakage current | ECG/bioimpedance electrodes, charger/debug interfaces | Define applied-part strategy; consider battery-only sensing and disable measurements during charging until proven safe. |
| Skin burn / thermal injury | LiPo pack, Qi coil, LTE bursts, display/backlight, regulators | Add NTC monitoring, thermal shutdown policy, charge-current limits, enclosure thermal validation. |
| Battery fire/swelling | LiPo charger/protection faults | Validate BQ25895, protection IC, NTC, current limits, short-circuit protection, UN38.3/IEC62133 cell documentation. |
| Measurement corruption | LTE/GNSS bursts, MIPI DSI, Qi magnetic field, switching regulators | Partition analogue/RF/power regions; add filtering, shielding, quiet rails, firmware state gating. |
| Overcurrent into body | AD5940 bioimpedance excitation fault | Set excitation limits, protection network, single-fault analysis, firmware limit checks. |
| False medical display/alert | Firmware, display, sensors, wireless link | Define claims, alarm behavior, verification plan, watchdog/failsafe strategy. |
Status
Primary Hazard Areas
Safety-Critical Subsystems
Required Decisions Before Schematic Freeze
Open Validation Items