3V3, GND, I2C_SDA, and I2C_SCL.HEATER_LOW_SIDE trace passing R10 with only about 0.36 mm local copper separation. It technically passes the present clearance rule, but it is poor margin in a noisy switching area.flux
Waiting for input
For Research Use Only. Not for use in diagnostic procedures.
Table
| Subsystem | Failure Mode | Effect | Severity (1–10) | Likelihood (1–10) | Detection (1–10) | RPN | Mitigation |
|---|---|---|---|---|---|---|---|
| Heater control | MOSFET stuck on | Overtemperature, tube damage, burn/fire risk | 8 | 3 | 4 | 96 | Independent series thermal cutoff; gate pulldown; watchdog; temperature bounds |
| Temperature sensing | Sensor open/short/detached | Incorrect control temperature | 8 | 4 | 3 | 96 | Plausibility diagnostics; heating-rate check; second independent cutoff |
| Firmware | Control loop freezes | Heater remains at last command | 8 | 3 | 4 | 96 | Hardware watchdog; heater-enable timeout; safe reset state |
| Power input | Wrong adapter voltage/polarity | Component damage or overheating | 7 | 3 | 3 | 63 | Keyed/labeled connector, reverse protection, TVS, fuse, input-voltage supervision |
| Thermal assembly | Insulation or mounting fails | Printed enclosure softens; accessible hot surface | 7 | 3 | 5 | 105 | Metal standoffs, thermal barrier, measured enclosure temperatures, guarded block |
| Optical chamber | Ambient light leak | Incorrect or unstable color measurement | 5 | 5 | 3 | 75 | LED-off dark check; lid switch; opaque liner; calibration limits |
| Tube placement | Tube not seated consistently | Temperature and optical errors | 6 | 5 | 5 | 150 | Keyed removable insert; positive seating feature; insertion check |
| Tube | Cap opens or tube leaks | Spill, contamination, invalid run | 7 | 2 | 6 | 84 | Use rated sealed tubes; containment recess; no user access during run |
| Assay interpretation | Sample pH/color mimics amplification | False-positive-looking result | 9 | 6 | 7 | 378 | No diagnostic call; record baseline/time series; validated controls and sample prep outside device |
| Assay interpretation | Inhibition or missing control | False-negative-looking result | 9 | 5 | 8 | 360 | Require external assay controls; report raw data/invalid conditions; no patient result |
| Carryover contamination | Amplified tube opened near preparation area | Future false positives | 8 | 4 | 8 | 256 | Closed-tube workflow; UDG-compatible chemistry; separated work areas; disposal procedure |
| User interface | Display implies clinical diagnosis | Unsafe reliance on unvalidated output | 10 | 3 | 7 | 210 | Show measurements and research status only; persistent RUO warning |
Status and Intended-Use Boundary
Future Regulatory Path
Biosafety Boundary
Selected Independent Thermal Cutoff
Preliminary FMEA
Priority Actions
Verification Required Before Energized Prototype Use
Regulatory Reference
Document Trigger