Wearable Health Patch
3D Model (STEP)
Failed to load 3D model
Table
| ID | Requirement | Current Implementation | Status |
|---|---|---|---|
| DI-001 | BLE wireless telemetry | MDBT50Q-1MV2 nRF52840 module | Draft |
| DI-002 | Battery-powered operation | LiPo connector + BQ24074 + buck-boost rails | Draft |
| DI-003 | USB charging | USB-C sink + BQ24074 | Draft |
| DI-004 | ECG acquisition | ADS1292R + FPC electrodes + protection | Draft |
| DI-005 | PPG heart-rate acquisition | MAX30102 | Draft |
| DI-006 | Lactate/glucose electrochemical acquisition | AD5940 + FPC electrodes | Research only |
| DI-007 | Contact/pressure sensing | Honeywell MPRLS0300YG00001B | Draft |
| DI-008 | Motion artifact data | KX022-1020 | Draft |
| DI-009 | Patient safety during USB | Analog switches + firmware control; hardware USB-force-disconnect still open | Incomplete |
| DI-010 | Battery safety | Charger TS net provisioned; battery protection/NTC finalization open | Incomplete |
Table
| Requirement | Verification Method | Status |
|---|---|---|
| Power rails | Bench bring-up, DMM/scope ripple | Planned |
| BLE telemetry | SWD flash + BLE advertising + data notify | Planned |
| ECG | ECG simulator + lead-off/contact tests | Planned |
| PPG | Optical fixture + motion/pressure tests | Planned |
| Electrochemical | Dummy cell + calibrated chemistry samples | Planned |
| Pressure/contact | Pressure source/calibrator | Planned |
| Battery charging | USB current limit, thermal rise, charge profile | Planned |
| USB patient safety | Leakage/current path test in USB attached and fault states | Planned |
Table
| Requirement | Schematic Output | Verification | Status |
|---|---|---|---|
| DI-001 | U1, J5, BLE antenna placement constraint | Firmware BLE test | Draft |
| DI-002 | J2, U2, U3, U4, C/L/R support | Power rail test | Draft |
| DI-004 | U5, J3, R16–R18, D3, U11 | ECG simulator | Draft |
| DI-006 | U7, J4, R19–R24, D3, U12/U13 | Dummy cell/calibration | Research |
| DI-009 | U11–U13, PATIENT_SW_CTRL | USB attached safety test | Incomplete |
Status
Applicable Standards / Frameworks to Evaluate
Device Classification
Intended Use
Design Inputs
Risk Analysis Summary
Design Outputs
Verification & Validation Plan
Traceability Matrix
Open Items
2026-07-27 Design Decision Record
2026-07-27 Layout Decision Record Update