O U LL Pinout and Electrical-Characterization Worksheet
Purpose
Determine the exact pin functions, connector orientation, voltage, current, load type, grounding, and isolation classification for legacy ports:
O: 2-position AC line-voltage connection
U: 2-position high-current auxiliary connection; function unknown
LL: 3-position auxiliary power/control connection; function unknown
No schematic connection, adapter wiring, or energized prototype connection is authorized until this worksheet is completed and independently reviewed.
Mandatory safety conditions
Treat every unknown pin as hazardous until proven otherwise.
Testing of energized or potentially mains-connected ports must be performed by a qualified person.
Use appropriately CAT-rated true-RMS instruments, insulated probes, PPE, barriers, strain relief, and an emergency disconnect.
Do not use continuity or resistance mode on an energized connector.
Before resistance testing, disconnect AC mains, batteries, backup supplies, capacitors, and every external cable that could backfeed the equipment.
Verify zero AC and DC voltage before touching conductors or changing meter modes.
Never measure prospective fault current by intentionally shorting pins.
Do not connect an earth-referenced oscilloscope to an unknown floating or mains-referenced circuit.
Stop testing if connector numbering, protective-earth relation, source isolation, or safe access is uncertain.
Port O must remain excluded from the low-voltage Bed Control Bench Prototype. If required, it belongs only in a separately enclosed hazardous-voltage interface.
Measure current only with a safe, correctly inserted instrument or isolated current probe.
Table
Pin/path
Idle current
Active current
Peak/inrush
Duty cycle
Signal waveform/frequency
Notes
LL1
LL2
LL3
LL1–LL2 path
LL1–LL3 path
LL2–LL3 path
LL functional investigation
Table
LL position
Destination/component
Proposed function
Source/load/control
Normal state
Confidence
1
2
3
Check all applicable classifications:
Supply positive/line
Supply return/neutral
Protective earth/shield
Switched output
Dry-contact common
Control input
Analog signal
Digital signal
Motor/actuator conductor
Unknown
LL derived assignment
Table
LL position
Final proposed name
Direction
Nominal voltage
Maximum current
Isolation domain
Status
1
[ ] Input [ ] Output [ ] Bidirectional
[ ] Verified [ ] TBD
2
[ ] Input [ ] Output [ ] Bidirectional
[ ] Verified [ ] TBD
3
[ ] Input [ ] Output [ ] Bidirectional
[ ] Verified [ ] TBD
LL disposition:
Safe for low-voltage controller PCB
Belongs only on separate hazardous/high-current module
Mechanical NC/DNP placeholder only
Exclude entirely
Blocked pending further testing
Isolation and domain classification
Complete after individual-port measurements.
Table
Boundary
Measured resistance
Measured working voltage
Required insulation type
Proposed project/domain
Pass/fail
O ↔ logic/SELV
[ ] Functional [ ] Basic [ ] Reinforced
U ↔ logic/SELV
[ ] Functional [ ] Basic [ ] Reinforced
LL ↔ logic/SELV
[ ] Functional [ ] Basic [ ] Reinforced
O ↔ chassis/earth
U ↔ chassis/earth
LL ↔ chassis/earth
O ↔ existing 120 VAC N/G/CC interface
U ↔ existing 120 VAC N/G/CC interface
LL ↔ existing 120 VAC N/G/CC interface
Insulation-resistance testing
Do not perform megohmmeter or hipot testing until all connected electronics are reviewed for test-voltage tolerance and isolated/disconnected as required.