The connectivity comparison confirms that all 49 nets still connect the same pins. The readability check found more work in the op-amp stages and electrode-jack wiring, which I’m correcting now.
The analog stages now have drawn signal paths, but a wire-placement bug on flipped symbols is blocking part of the USB drawing. I’ll leave the affected connections labeled and finish the independent label fixes; the bug needs Contact Support.
I arranged all 78 components into eight functional blocks and added local wires and net labels. All 49 nets retain exactly the same pin connections; no electrical connections changed.
The drawing is not fully clean: wire crossings and label crowding remain, and a flipped-symbol wire-placement bug blocks the shared USB power wire. That bug needs Contact Support; the remaining drafting cleanup is still open.
Current implemented AD8232ACPZ-RL / MCP6002-E/SN circuit: 71 components / 44 connected nets. Native ngspice AC/DC/linear transient runs are distinct from analytical calculations. No official nonlinear AD8232 model was executed. Complete values, equations, source conditions and raw evidence are in Simulation Guide.
The native signal model uses actual target passives, assumed 2 pF clamp-midpoint capacitance, ideal AD gain 100 and servo with 8.06 ms effective time constant, postgain 1 / 5.02 with typical-GBW pole, MCP one-pole 1 MHz followers with 6 pF grounded / 3 pF bootstrapped input capacitance, and the actual 100 Ω / 1 nF ADC interface. It omits nonlinear clamps, clipping, real servo range, RLD, reference dynamics, noise and ADC switching. There is no extra hidden INA pole in the principal run.
Native AC
Table
Metric
J4 open
J4 shunted
Gain at 20 Hz
68.69158
344.8316
Gain at 100 Hz
94.82377
476.0096
Gain at 500 Hz
67.75089
340.0066
Peak V/V
96.58724
484.8338
−3.000 dB edges, Hz
19.8167–497.715
19.8139–497.658
Half-power edges, Hz
19.7696–498.011
19.7669–497.954
4 kHz rejection versus 100 Hz
74.6568 dB
74.8214 dB
7.5 kHz rejection
99.8257 dB
100.377 dB
8.02 kHz rejection
102.600 dB
103.224 dB
Density refinement from 400 to 800 points/decade changed edges ≤0.003 Hz. Peak-frequency reports differ by approximately one grid interval; no optimized peak frequency is claimed. LP2_A maximum is 98.70065 / 495.4263 V/V, above ADC peak, so every node matters for headroom.
Native DC and transient demand
Ideal DC cancellation at jack offsets ±300 mV gives IAOUT about 1.6 V while numerical cancellation state reaches ±28.9575 V: not physical internal-range validation. Conditional +4.5 mV offsets per MCP follower add about 9 mV ADC baseline.
Pulse: 1 µs edges, onset 20 ms, hold 100 ms, record 200 ms. The 1/10 mV results are exact scaling of the executed 100 mV linear response:
Table
Jack step
INA_OUT peak excursion
ADC open
ADC shunted
1 mV
0.094196 V
0.097408 V
0.488964 V
10 mV
0.941960 V
0.974076 V
4.889641 V
100 mV
9.419600 V
9.740758 V
48.896410 V
Impossible rail demands are intentional model outputs, not physical clipping/recovery predictions. Maximum-timestep refinement from 10 to 5 µs changed positive peaks ≤0.0028%, all extrema <0.0037%, screen durations ≤8.1 µs. Do not derive a firmware discard interval from these demands.
Independent calculations
512 deterministic passive endpoint cases, not Monte Carlo/yield: lower edge 17.78–23.99 Hz, upper edge 476.79–537.65 Hz, peak 93.91–102.79 V/V for the documented ideal-follower case. No exact 500 Hz production mask.
Source-specific finite-Cin partial noise: 3.21181 / 3.21085 µVrms at 300 K, integrated 20–500 Hz, jack-referred using 100 Hz gain. Important internal/current/reference/servo/electrode/ADC sources are missing, not zero. Integration refinement agrees within 0.000013%.
Ideal 12-bit quantization at 3.2 V: 781.25 µV/LSB, 78.125 µV RMS in 20–500 Hz at 8 kS/s. Partial SNR for 10/100/1000 µVrms, 100 Hz sine: 9.59/29.59/49.59 dB open and 9.86/29.86/49.86 dB shunted. Not measured ENOB or complete-board SNR.
Source-specific downstream modeled white-noise aliases add 9.5209 / 9.6140 / 9.6290 µV ADC RMS at 400/800/1600 images; final change 0.156%. Extrapolated spectra and ideal point sampling remain limitations.
Separate 648-case three-branch RLD plant with C30 = 10 nF: crossover 63.79–94.35 Hz, minimum PM 55.33°, modeled 60 Hz benefit 0.44–5.23 dB. With C tolerance and an assumed 1 kHz active pole, minimum PM is 50.19°; real-loop stability is unvalidated.