Regulatory and Safety Notes — Wearable EEG BCI Acquisition Board
Status
Draft design-stage notes for a non-medical human-connected EEG/BCI prototype.
Intended Use
  • Non-medical research/prototype use only.
  • Not intended for diagnosis, treatment, patient monitoring, or life-sustaining operation.
  • Human-subject testing requires institutional review, informed consent, and an independent safety review.
Wireless / EMC Notes
  • U3 ESP32-WROOM-32E provides Wi-Fi/Bluetooth radio functionality.
  • Preserve module integration requirements: antenna edge placement, no copper under antenna keepout, no shielding/enclosure metal over antenna, and follow Espressif certification conditions.
  • Likely documentation needed for a productized device: FCC Part 15/CE RED technical file, BOM, schematic, PCB layout, module datasheet/certification, user manual, and lab test reports.
Battery and Charging Safety
  • U4 MCP73831 charges a single-cell LiPo from VBUS.
  • J5 LiPo polarity must be mechanically/keying verified.
  • If the chosen LiPo cell lacks protection, add protected-cell requirement or include protection/fuel-gauge circuitry in the design.
  • Verify charger thermal performance and charge current before enclosing the wearable.
Human-Connected Electrode Safety

Table


HazardSourceCurrent MitigationOpen Action
Excess electrode currentPower fault or BIAS drive fault4.7k series resistors per channel, ESD clamps to AGND, 1k BIAS series resistorIndependent leakage/current-limit analysis required
ESD into AFEElectrode handlingESD601 clamps to AGND on CH1–CH8Confirm clamp capacitance/leakage is acceptable for EEG bandwidth
Ground/USB coupling to userUSB-connected PC while electrodes attachedNone beyond circuit-level groundingConsider battery-only operation during human recording or medically isolated USB
RF/switching noise in EEG bandESP32 and switchers near AFELayout zoning plannedValidate with noise measurements and shielding strategy
Incorrect electrode pinoutJ3 cable mismatchJ3 pinout documentedConfirm mating cable before use
Preliminary FMEA

Table


SubsystemFailure ModeEffectSeverityMitigation
LiPo/chargerOvercharge/thermal faultBattery damage, heat8Use protected cells, verify MCP73831 thermal design
Power tree+5V0A over/undervoltageADS1299 malfunction6Bring-up rail checks, LDO validation
Electrode inputESD surgeAFE damage, false readings6Series resistors and ESD clamps
Bias driveIncorrect BIASREF/BIASINV/BIASIN topologyBad common-mode control or unsafe electrode drive8Finalize ADS1299 bias circuit before human testing
USB connectionNon-isolated PC ground couplingUser leakage/ground loop risk9Prefer battery-only acquisition or isolated USB for human tests
ESP32 RFAntenna keepout violationPoor radio/EMC failures5Edge placement and no-copper keepout
Productization Open Items
  • Decide whether this remains a lab prototype or becomes a regulated product.
  • If productized, prepare full FCC/CE documentation and safety risk file.
  • For any medical claims or clinical use, create a formal design history file and follow applicable medical-device standards; current design is not sufficient for that use.