| Hazard | Source | Current Mitigation | Open Action |
|---|---|---|---|
| Excess electrode current | Power fault or BIAS drive fault | 4.7k series resistors per channel, ESD clamps to AGND, 1k BIAS series resistor | Independent leakage/current-limit analysis required |
| ESD into AFE | Electrode handling | ESD601 clamps to AGND on CH1–CH8 | Confirm clamp capacitance/leakage is acceptable for EEG bandwidth |
| Ground/USB coupling to user | USB-connected PC while electrodes attached | None beyond circuit-level grounding | Consider battery-only operation during human recording or medically isolated USB |
| RF/switching noise in EEG band | ESP32 and switchers near AFE | Layout zoning planned | Validate with noise measurements and shielding strategy |
| Incorrect electrode pinout | J3 cable mismatch | J3 pinout documented | Confirm mating cable before use |
Table
| Subsystem | Failure Mode | Effect | Severity | Mitigation |
|---|---|---|---|---|
| LiPo/charger | Overcharge/thermal fault | Battery damage, heat | 8 | Use protected cells, verify MCP73831 thermal design |
| Power tree | +5V0A over/undervoltage | ADS1299 malfunction | 6 | Bring-up rail checks, LDO validation |
| Electrode input | ESD surge | AFE damage, false readings | 6 | Series resistors and ESD clamps |
| Bias drive | Incorrect BIASREF/BIASINV/BIASIN topology | Bad common-mode control or unsafe electrode drive | 8 | Finalize ADS1299 bias circuit before human testing |
| USB connection | Non-isolated PC ground coupling | User leakage/ground loop risk | 9 | Prefer battery-only acquisition or isolated USB for human tests |
| ESP32 RF | Antenna keepout violation | Poor radio/EMC failures | 5 | Edge placement and no-copper keepout |
Status
Intended Use
Wireless / EMC Notes
Battery and Charging Safety
Human-Connected Electrode Safety
Preliminary FMEA
Productization Open Items