EOG Front-End Requirements and Handoff
Requested deliverable
Professional schematic suitable for PCB implementation for a battery-powered differential EOG eye movement and blink detector. User requests a medical-grade design; medical safety, compliance, and clinical suitability have not been established.
User requirements
- 3.7 V Li-ion battery; portable device.
- STM32F401 ADC interface.
- Differential EOG measurement.
- Total gain approximately 1000 V/V.
- Passband target 0.05 Hz–30 Hz.
- Virtual ground at 1.65 V.
- INA333 instrumentation amplifier.
- High-pass filtering for baseline drift removal.
- Second-order active low-pass filter.
- Electrode protection resistors, low-noise implementation, and ADC anti-alias filtering.
- Clearly labeled schematic stages: (1) Input protection, (2) Instrumentation amplifier, (3) High-pass filter, (4) Gain stage, (5) Low-pass filter, (6) STM32 ADC interface.
Actual session state
Only this requirements/handoff document was created. No parts, nets, schematic wiring, PCB layout, simulation, datasheet research, or electrical verification were completed. The starting project had zero components and zero nets. Execution stopped because the account ran out of ACUs.
Decisions and evidence still required
- Define intended use, applicable medical safety requirements, electrode arrangement, reference/bias electrode, expected differential signal and electrode DC offsets, electrode impedance, and input common-mode range.
- Define whether USB, charging, programming, or any external wired equipment can be connected while electrodes contact a person. Battery operation alone does not establish patient isolation or safety.
- Establish patient current limits and protection under normal and single-fault conditions before selecting protection parts; protection resistors alone do not establish medical compliance.
- Confirm regulated analog/ADC supply and reference arrangements. A 3.7 V nominal Li-ion cell is not a regulated 3.3 V rail; verify full battery voltage range against all parts.
- Verify INA333 supply, input common-mode limits, output swing, noise, bias-current behavior, and recovery using its datasheet.
- Allocate gain so electrode DC offset cannot saturate the instrumentation amplifier before the high-pass stage. Do not place the full 1000 V/V gain ahead of baseline removal without headroom analysis.
- Select and verify rail-to-rail low-noise amplifiers, buffered 1.65 V reference, high-pass topology and leakage-sensitive components, low-pass response/Q, and ADC drive circuit.
- Specify sampling rate and required alias rejection. A 30 Hz second-order low-pass alone does not establish adequate anti-aliasing at an unspecified sample rate.
- Verify ADC sample-and-hold settling, allowable source impedance, output excursion, noise budget, tolerances, startup/overload recovery, and power consumption.
Next implementation work
Research verified datasheets and patient-interface constraints, resolve safety-critical missing requirements, then build and label the six stages. Verify connectivity/ERC, gain and frequency response, headroom with electrode offsets, noise, ADC settling, and protection behavior before presenting a PCB-ready schematic. Treat any initial implementation as an unvalidated prototype, not a certified medical device. Do not test on people before appropriate safety review and verification.