Purpose
Design a complete integrated wearable ultrasound system schematic for a wearable therapy and sensing device with two operating modes: A-mode sensing and LIPUS therapy.
Primary transducer
Single 3 MHz, 10 mm piezoelectric transducer
Operating modes
A-mode sensing: short pulse transmit, protected receive path, analog front end amplification/filtering, digitization, and MCU processing
LIPUS therapy: controlled therapeutic burst drive with safety interlocks and continuous monitoring
Architecture
Diagram
Block Requirements
Power management block with LiPo battery input, protected charging and power path, 3.3 V logic rail, regulated +15 V analog rail, regulated -15 V analog rail, and a higher-voltage pulser rail
Microcontroller block for mode selection, pulse timing, burst control, safety interlock enforcement, and ADC data handling
Pulser block for generating short transmit pulses and gated therapy bursts centered at 3 MHz
T/R protection block that isolates the receive chain during transmit and clamps transducer energy away from the receive path
Analog front end block including receive protection, variable gain, post-amplification, envelope detection, and band-limiting around 3 MHz
ADC conditioning block for scaling and filtering the detected signal into the MCU ADC range
Safety sensor block including at minimum temperature, current, battery or supply status, and audible/visual user alerts under MCU supervision
Interlock Strategy
Therapy mode is enabled only when no safety fault is active
Any overtemperature, overcurrent, low battery, or poor contact condition disables the pulser path
MCU asserts mutually exclusive control of sensing and therapy states
Receive chain is disabled or protected during transmit events
Implementation Notes
Use clear named nets for rails, pulser control, T/R control, ADC interface, sensor alarms, and transducer nodes
Use an STM32 control block with SWD access and a DS18B20-compatible GPIO plus local clock support
Keep mixed-signal grounding disciplined, with the analog receive path referenced cleanly and the digital controller return separated in routing intent until the ADC interface region
Review the finished schematic for complete block interconnection and interlock coverage
Design Summary
Status: Draft
Manufacturing target: Prototype
Software / firmware: STM32 firmware for pulse timing, receive acquisition, safety interlocks, and user alert handling
Scope
Purpose
This board integrates battery charging, rail generation, ultrasound transmit drive, protected receive amplification, signal detection, MCU control, and safety monitoring into one wearable ultrasound electronics platform.
In scope
Single-cell LiPo input, charger, and system power path
3.3 V, +15 V, -15 V, and pulser high-voltage generation
Temperature, current, and battery-status monitoring with LED and buzzer indicators
Out of scope
Final enclosure mechanics
Final therapy waveform validation in simulation or bench characterization
System context
The board is the central electronics engine of a wearable ultrasound treatment and sensing device. It powers a single-element transducer, acquires response data, enforces treatment safety, and exposes debug access for firmware development.
Requirements
Functional
The board shall drive a single 3 MHz ultrasound transducer for transmit and therapy modes.
The board shall acquire the protected receive signal and present a conditioned detection signal to the MCU ADC.
The board shall monitor battery, temperature, and pulser current and shall disable therapy on fault.
The board shall provide LED and buzzer indication for fault or operating status.
Electrical
Input power: single-cell LiPo with USB charging and system power-path support
MCU timing target: 3 MHz pulse and timing generation capability
Receive chain target: 40 dB to 50 dB total gain with bandwidth above 3 MHz
Critical interfaces: SWD, I2C fuel gauge and charger telemetry, ADC input conditioning, transducer interface, LED and buzzer control
Mechanical / environmental
Wearable use implies compact PCB area, low-profile assembly, and thermally aware placement of charger and power converters away from sensing points
Thermal safety monitoring is required during therapy operation
Key constraints
Use production-realistic ICs already present in the design where feasible
Preserve SWD debug access and named nets for all major rails and controls
Keep the analog receive path and its grounding strategy distinct from noisy switching and digital timing domains
Include sufficient local 100 nF decoupling plus bulk rail capacitance at each powered subsystem
Dependencies and risks
Dependencies
Firmware support for pulse sequencing, ADC acquisition, safety state machine, and user feedback control
Bench characterization of pulser amplitude, receive gain, and transducer matching
Key risks
Placeholder or partially connected blocks can hide unsafe therapy behavior until fully closed out
Receive path overload during transmit can saturate or damage the analog front end if T/R protection is incomplete
Mixed-signal ground return errors can degrade the envelope and ADC signal quality
Validation
Success criteria
All charger and rail-generation blocks are present and coherently connected
MCU, control nets, and safety interfaces are wired for functional firmware control
Signal flow exists from transducer connector through protection, receive gain, envelope detection, and ADC conditioning to the MCU ADC
Faults can inhibit the pulser path and drive user-visible alerts
Planned checks
ERC review for floating pins and single-ended dead nets
Datasheet cross-check for power, ADC, and analog front-end pin usage
Bench validation of rail voltages, pulser current sense, over-temperature shutdown, and ADC envelope scaling
Release-facing notes
Expected deliverables
Integrated schematic
BOM with key ICs and generic passives
Bring-up and safety verification notes
Special release notes
Keep test access to major rails and debug signals
Treat pulser, protection, and receive sections as safety-critical during review
Change notes / open questions
Changes in this revision
Captured the integrated wearable ultrasound architecture and subsystem requirements
Established rail, signal-flow, and safety-interlock intent for schematic completion
Open questions
Final device choices for negative rail generation, final pulser stage topology, and final indicator implementation will be resolved during schematic completion